Back to Search
Start Over
Sensitive immunoassay of cardiac troponin I using an optimized microelectrode array in a novel integrated microfluidic electrochemical device
- Source :
- Analytical and Bioanalytical Chemistry. 412:8325-8338
- Publication Year :
- 2020
- Publisher :
- Springer Science and Business Media LLC, 2020.
-
Abstract
- A sensitive and portable microfluidic electrochemical array device (μFED) was developed for the immunoassay of trace amounts of human cardiac troponin I (cTnI), which is an attractive biomarker for acute myocardial infarction (AMI). The classical "sandwich" method was adopted for the immunoassay. The capture antibody was immobilized using a self-assembled monolayer (SAM) technique, and the process was reorganized to be compatible with the bonding process. The detection antibody was labeled with alkaline phosphatase (AP) for signal amplification. The performance of the μFED was improved by eliminating the shielding effect of the microelectrode array (MEA) integrated in the μFED. The effects of the interstice and the width of the MEA on the response peak current were analyzed and simulated. The concentration gradient, about 3% of the gradient at the surface, was considered as the criterion for estimation of the optimal interstice between electrodes, and its effectiveness was proved. A stable and miniaturized reference electrode was integrated in the μFED, and its potential deviation was less than 5 mV in 15 min. These efforts resulted in the enhanced immunoassay performance of the μFED. A low limit of detection of about 5 pg/mL was obtained in serum samples, and the response current was proportional to the logarithm of concentration from 50 pg/mL to 1 μg/mL. The immunoassay process was accomplished in 15 min. The μFED was thus qualified and is a promising candidate for point-of-care immunoassay of cTnI. Graphical abstract.
- Subjects :
- Materials science
Point-of-Care Systems
Microfluidics
02 engineering and technology
Electrochemistry
01 natural sciences
Biochemistry
Reference electrode
Analytical Chemistry
Limit of Detection
Lab-On-A-Chip Devices
Troponin I
medicine
Humans
Immunoassay
Detection limit
Chromatography
medicine.diagnostic_test
010401 analytical chemistry
Electrochemical Techniques
Multielectrode array
021001 nanoscience & nanotechnology
0104 chemical sciences
Electrode
0210 nano-technology
Microelectrodes
Biomarkers
Subjects
Details
- ISSN :
- 16182650 and 16182642
- Volume :
- 412
- Database :
- OpenAIRE
- Journal :
- Analytical and Bioanalytical Chemistry
- Accession number :
- edsair.doi.dedup.....bbab6cb2e9b766a0e7d01b9709b75f4e
- Full Text :
- https://doi.org/10.1007/s00216-020-02968-z