CT Body Stereotaxic Instrument for Percutaneous Biopsy and Other Interventive Procedures Phantom Studies
- Source :
- Investigative Radiology. 20:525-526
- Publication Year :
- 1985
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 1985.
-
Abstract
- This article describes a new body stereotaxic system that defines a fiducial point by means of a skin localization device placed directly on the patient. The system uses a rectilinear stereotaxic frame to guide the needles along the calculated path. A method for calculating paths that require angulations from one scan slice to another is described, as well. The system was tested in a foam phantom and shown to be intrinsically accurate in vitro to within 2 mm in both the x and y axis.