Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula

Authors :
Casper F Zijderhand
Wiebe G Knol
Ricardo P J Budde
Cornelis W van der Heiden
Kevin M Veen
Jelena Sjatskig
Olivier C Manintveld
Alina A Constantinescu
Ozcan Birim
Jos A Bekkers
Ad J J C Bogers
Kadir Caliskan
Cardiothoracic Surgery
Radiology & Nuclear Medicine
Cardiology
Source :
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery, 62(4):ezac415. Elsevier
Publication Year :
2022

Abstract

OBJECTIVES This retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms. METHODS Medical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized protocol was created to measure the angular position with a contrast-enhanced computed tomography scan. Statistics were done using a gamma frailty model with a constant rate function. RESULTS For this analysis, 48 LVAD-supported patients were included. The majority of the patients were male (79%) with a median age of 57 years and a median follow-up of 30 months (interquartile range: 19–41). Low-flow alarm(s) were experienced in 30 (63%) patients. Angulation towards the septal–lateral plane showed a significant increase in low-flow alarms over time with a constant rate function of 0.031 increase in low-flow alarms per month of follow-up per increasing degree of angulation (P = 0.048). When dividing this group using an optimal cut-off point, a significant increase in low-flow alarms was observed when the septal–lateral angulation was 28° or more (P = 0.001). Anterior–posterior and maximal inflow cannula angulation did not show a significant difference. CONCLUSIONS This study showed an increasing number of low-flow alarms when the degrees of LVAD inflow cannula expand towards the septal–lateral plane. This emphasizes the importance of the LVAD inflow cannula angular position to prevent relapsing low-flow alarms with the risk of diminished quality of life and morbidity.

Details

ISSN :
1873734X and 10107940
Volume :
62
Issue :
4
Database :
OpenAIRE
Journal :
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
Accession number :
edsair.doi.dedup.....ba76c67d847268ce8c955cd2cde48eb6