A Graphenic Biosensor for Real-Time Monitoring of Urea During Dialysis
- Source :
- IEEE Sensors Journal. 20:4571-4578
- Publication Year :
- 2020
- Publisher :
- Institute of Electrical and Electronics Engineers (IEEE), 2020.
-
Abstract
- A patient at the fifth stage of chronic kidney disease usually needs dialysis or transplantation. The adequacy of dialysis can be assessed by the dimensionless parameter Kt/V , which depends on the concentration of urea in the blood. Kt/V is usually measured before and after treatment, or by an indirect measurement of dialysis fluid conductivity. For the real-time monitoring of the urea concentration during dialysis, this paper proposes a disposable biosensor that combines a pH sensitive reduced graphene oxide layer functionalized with 4-amino benzoic acid and urease. Urease is immobilized on reduced graphene oxide and catalyzes the hydrolysis of urea into carbon dioxide and ammonia. This catalysis increases the pH value locally and elicits a sensor response. Urea concentration was assessed in the plasma of five dialyzed patients using a potentiometric measurement. The response time was 120 s and the error was 6 ± 3% compared with standard clinical laboratory analysis.
- Subjects :
- Urease
Potentiometric titration
urea
reduced graphene oxide
01 natural sciences
Ammonia
chemistry.chemical_compound
Electrical and Electronic Engineering
Instrumentation
Benzoic acid
urease
Chromatography
biology
pH
010401 analytical chemistry
chronic kidney failure
Enzymatic biosensor, reduced graphene oxide, 4-amino benzoic acid, urease, pH, chronic kidney failure, dialysis, urea
4-amino benzoic acid
0104 chemical sciences
Transplantation
chemistry
biology.protein
Urea
Enzymatic biosensor
dialysis
Dialysis (biochemistry)
Biosensor
Subjects
Details
- ISSN :
- 23799153 and 1530437X
- Volume :
- 20
- Database :
- OpenAIRE
- Journal :
- IEEE Sensors Journal
- Accession number :
- edsair.doi.dedup.....ba56c7c04ed972abf94300b4b9c2f673