Back to Search Start Over

Magnetic resonance imaging with stepped B0 fields

Authors :
Martyn N.J. Paley
Paul D. Griffiths
K.J. Lee
Iain D. Wilkinson
Source :
Magnetic Resonance Imaging. 21:625-629
Publication Year :
2003
Publisher :
Elsevier BV, 2003.

Abstract

This paper introduces the concept of imaging with stepped magnetic fields within the main B 0 field. The resonance frequency within each step is separated from the next by an amount greater than the imaging bandwidth. Each step thus gives rise to a separate imaging region enabling simultaneous imaging over multiple slices without aliasing. The auxiliary magnetic field regions are produced by coil inserts. The principle is illustrated using simple coil designs which introduce up to two steps into the main B 0 field. Phantoms placed in each uniform region were successfully imaged simultaneously with no aliasing.