The effect of measurement area size on the reliability of myocardial iron load measurement in cardiac magnetic resonance imaging examinations of thalassemia patients

Authors :
Fatih Mehmet Azik
Funda Dinç Elibol
İbrahim Önder Yeniçeri
Source :
Turkish Journal of Medical Sciences
Publication Year :
2021
Publisher :
The Scientific and Technological Research Council of Turkey, 2021.

Abstract

Background/aim The aim of this study was to evaluate the intraobserver and interobserver reliability of cardiac T2* MRI measurements in different region of interest (ROI) sizes. Materials and methods Cardiac T2* MRIs of 24 thalassemia major patients were evaluated. Two different ROI sizes were used for measurement. In the first measurement, an ROI approximately 5 mm in diameter was used in the interventricular septal myocardium. In the other method, the whole ventricular septal myocardium was used as the measurement. The intraobserver and interobserver variabilities were assessed with the intraclass correlation coefficient (ICC). Results The measurement of the first observer, the ICC of the small-sized ROI (ssROI), was 0.869, and the measurement for the second observer, the ICC of the ssROI, was 0.659. The ICC of the whole-septal ROI (wsROI) was 0.991 for the first observer and 0.980 for the second observer. Interobserver variability, for the mean measurement, was 0.442 for the ICC of ssROI and 0.883 for the ICC of wsROI. Conclusion For the evaluation of myocardial iron load with T2* MRI we suggest making measurements with ROI, including all of the interventricular septum, as a consequence of high intraobserver and interobserver consistency.

Details

Language :
English
ISSN :
13036165 and 13000144
Volume :
51
Issue :
2
Database :
OpenAIRE
Journal :
Turkish Journal of Medical Sciences
Accession number :
edsair.doi.dedup.....b51fe2715112c268d2920b15a50e848b