HeartWare HVAD Flow Estimator Accuracy for Left and Right Ventricular Support

Authors :
Shaun D. Gregory
Eric L. Wu
Andrew Stephens
Ulrich Steinseifer
John F. Fraser
Nicole Bartnikowski
Martin Mapley
Source :
ASAIO journal (American Society for Artificial Internal Organs : 1992). 67(4)
Publication Year :
2021

Abstract

This study investigated the accuracy of the HeartWare HVAD flow estimator for left ventricular assist device (LVAD) support and biventricular assist device (BiVAD) support for modes of reduced speed (BiVAD-RS) and banded outflow (BiVAD-B). The HVAD flow estimator was evaluated in a mock circulatory loop under changes in systemic and pulmonary vascular resistance, heart rate, central venous pressure, and simulated hematocrit (correlated to viscosity). A difference was found between mean estimated and mean measured flow for LVAD (0.1 ± 0.3 L/min), BiVAD-RS (-0.1 ± 0.2 L/min), and BiVAD-B (0 ± 0.2 L/min). Analysis of the flow waveform pulsatility showed good correlation for LVAD (r2 = 0.98) with a modest spread in error (0.7 ± 0.1 L/min), while BiVAD-RS and BiVAD-B showed similar spread in error (0.7 ± 0.3 and 0.7 ± 0.2 L/min, respectively), with much lower correlation (r2 = 0.85 and r2 = 0.60, respectively). This study demonstrated that the mean flow error of the HVAD flow estimator is similar when the device is used in LVAD, BiVAD-RS, or BiVAD-B configuration. However, the instantaneous flow waveform should be interpreted with caution, particularly in the cases of BiVAD support.

Details

ISSN :
1538943X
Volume :
67
Issue :
4
Database :
OpenAIRE
Journal :
ASAIO journal (American Society for Artificial Internal Organs : 1992)
Accession number :
edsair.doi.dedup.....b41c41dcaad1111e73023deb08918ba4