Back to Search Start Over

Using negative bolus in dynamic MPI

Authors :
Mohn, Fabian
Exner, Miriam
Szwargulski, Patryk
Möddel, Martin
Knopp, Tobias
Gräser, Matthias
Source :
International Journal on Magnetic Particle Imaging 9 (1): 2303022 (2023)
Publication Year :
2023

Abstract

In Magnetic Particle Imaging, the spatial distribution of a tracer is measured and depicted with a concentration dependent signal intensity for any location that inhibits particles, whereas surrounding tissue does not provide any signal. After tracer injection, the signal over time (positive contrast) can be utilized as a transient response to calculate dynamic diagnostic parameters like perfusion parameter maps. In this work, a bolus of physiological saline solution without any particles (negative contrast) is proposed, where the remaining steady state concentration contributes to the image contrast. This opens up the possibility to stretch the total monitoring time of a patient by utilizing a positive-negative contrast sequence, while keeping the total iron dose constant in the subject. Resulting time responses show that normalized signals from positive and negative boli are concurrent in the phantom experiments, indicating identical diagnostic parameters for in-vivo use.<br />International Journal on Magnetic Particle Imaging IJMPI, Vol 9 No 1 Suppl 1 (2023)

Details

ISSN :
23659033
Database :
OpenAIRE
Journal :
International journal on magnetic particle imaging
Accession number :