Back to Search Start Over

The effect of glaucoma treatment using high-intensity focused ultrasound on total and corneal astigmatism: a prospective multicentre study

Authors :
Michele Figus
Alix Somers
S. Melamed
David Cordeiro Sousa
Luís Abegão Pinto
Raquel Esteves Marques
Evelien Vandewalle
Marco Nardi
Ingeborg Stalmans
Publication Year :
2020

Abstract

PURPOSE Ultrasound cycloplasty (UCP) acts through the selective coagulation of the ciliary body using high-intensity focused ultrasound (HIFU) technology. The aim of this study was to investigate whether the application of ultrasound beams targeting the ciliary body using an external probe influences astigmatism. METHODS Multicentre, prospective, single-arm, open-label study in adult patients with primary open-angle glaucoma and moderately uncontrolled intraocular pressure (IOP) under glaucoma medication. The primary outcome was induced corneal astigmatism, calculated from topography, and assessed statistically through vector analysis. Secondary outcomes included induced total astigmatism and mean changes from baseline in best-corrected visual acuity (logMAR) and IOP. Subgroup analysis was performed to assess the impact of device centring on corneal and total induced astigmatism. RESULTS Fifty eyes were enrolled. Mean age was 69.6 ± 11.3 years. At 1, 3 and 6 months postprocedure, HIFU-induced corneal astigmatism was 0.88 D × 93°, 0.87 D × 106° and 1.16 D × 97°, respectively, while induced total astigmatism was 0.62 D × 103°, 0.42 × 106° and 0.39 × 107°. By the last follow-up, the percentage of patients with

Details

Language :
English
Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....b3cac4b1788e653d8b92cccfa0984f63