Abstract
- We studied the spherical aberration of an intraocular lens, using third-order theory, ray tracing, and bench measurements by the Ronchi test. Good agreement was obtained for these three methods of investigation. In particular we found that the Ronchi test is an accurate method for quantifying this aberration.
- Subjects :
- Wavefront
Materials science
Intraocular lens
Optical properties
business.industry
Materials Science (miscellaneous)
medicine.medical_treatment
Holographic interferometry
Industrial and Manufacturing Engineering
Ronchi test
Spherical aberration
Interferometry
Optics
Intraocular lenses
medicine
Ray tracing (graphics)
Business and International Management
business
Óptica
Optical aberration
Subjects
Details
- ISSN :
- 15394522 and 00036935
- Volume :
- 32
- Database :
- OpenAIRE
- Journal :
- Applied Optics
- Accession number :
- edsair.doi.dedup.....b336509fc5aa18b5f6ea80afc945427a
- Full Text :
- https://doi.org/10.1364/ao.32.004132