Back to Search
Start Over
Late Fiber Metal Shedding of the First and Second-Generation Harris Galante Acetabular Component. A Report of 5 Cases
- Source :
- The Journal of Arthroplasty. 22:624-629
- Publication Year :
- 2007
- Publisher :
- Elsevier BV, 2007.
-
Abstract
- Five patients presented with fiber metal mesh shedding of a Harris Galante II acetabular cup detected between 11 and 15 years after implantation. All patients presented with hip pain and 4 demonstrated gross acetabular loosening and fiber metal separation on preoperative radiographs. The remaining patient underwent revision surgery because of a liner dislodgment and had a radiographically well fixed shell. Loosening and fiber metal separation were detected intraoperatively. Scanning electron microscopy of the retrieved shells demonstrated isolated diffusion bonding marks in the areas where the mesh separated from the substrate and no evidence of corrosion. Progressive osteolysis in the iliac bone was evident in 4 of our cases. Progressive iliac osteolysis may lead to loss of bone support in well-fixed cups and excessive stresses transferred to the interface between the fiber metal mesh and the titanium substrate leading to the separation of the 2 layers. Fiber metal separation may contribute to long-term loosening in the Harris Galante acetabular component.
- Subjects :
- Adult
musculoskeletal diseases
medicine.medical_specialty
Time Factors
Osteolysis
Dentistry
Iliac bone
Humans
Medicine
Orthopedics and Sports Medicine
Hip pain
Fiber
Aged
business.industry
Acetabulum
Middle Aged
musculoskeletal system
equipment and supplies
medicine.disease
Prosthesis Failure
Surgery
Acetabular component
Disease Progression
Metal mesh
Female
Hip Prosthesis
business
Subjects
Details
- ISSN :
- 08835403
- Volume :
- 22
- Database :
- OpenAIRE
- Journal :
- The Journal of Arthroplasty
- Accession number :
- edsair.doi.dedup.....afa0cd4d6233c46b0cc07e8a24217371
- Full Text :
- https://doi.org/10.1016/j.arth.2006.05.027