Accuracy of an Automated Hearing Aid Fitting Using Real Ear Measures Embedded in a Manufacturer Fitting Software
- Source :
- Journal of the American Academy of Audiology. 32:157-163
- Publication Year :
- 2021
- Publisher :
- Georg Thieme Verlag KG, 2021.
-
Abstract
- Background Hearing aid fitting guidelines recommend real ear measures (REM) to verify hearing aid performance. Unfortunately, approximately 70 to 80% of clinicians do not use REM, but instead download manufacturer first-fit. Studies report differences in performance between first-fit and programmed-fit with greatest differences in the higher frequencies. Recently, hearing aid and real ear analyzer (REA) manufacturers allow REA communication with hearing aid software feature to automatically program hearing aids to target. Little research is available reporting the accuracy of this feature. Purpose The aim of the study is to examine whether differences exist at 50, 65, and 80 dB SPL between two ReSound first-fit formulae (Audiogram+ and NAL-NL2) using ReSound AutoREM and Aurical NAL-NL2 Research Design The study design is of repeated measure type. Study Sample The study sample includes 48 ears. Data Collection and Analysis For the two fitting formulae, AutoREM real ear insertion gain (REIG) was measured at 50, 65, and 80 dB SPL and compared with measures from Aurical NAL-NL2. Results Mean AutoREM REIG for ReSound NAL-NL2 was 3 to 8 dB below Aurical NAL-NL2 for 50 dB SPL, within 1 to 3 dB for 65 dB SPL and 1 to 5 dB above for 80 dB SPL. Mean AutoREM REIG for Audiogram + was 1 to 12 dB below Aurical NAL-NL2 for 50 dB SPL, within 2 to 5 dB for 65 dB SPL and 1 to 7 dB above NAL-NL2 for 80 dB SPL. Conclusion Relative to the Aurical NAL-NL2, AutoREM REIG50 for Audiogram + and ReSound NAL-NL2 was lower. Relative to the Aurical NAL-NL2, AutoREM REIG65 for Audiogram + was higher at 1,000 Hz and lower at 4,000 to 6,000 Hz and for ReSound NAL-NL2 it was lower at 500 Hz and 4,000 Hz and higher at 3,000 Hz. Relative to the Aurical NAL-NL2, AutoREM REIG80 for Audiogram + was higher at 500 to 3,000 Hz and 6,000 Hz and ReSound NAL-NL2 was higher at 500 to 6,000 Hz. Because of wide intersubject variability clinicians should continue to use REM as a “check and balance” when using AutoREM.
- Subjects :
- Hearing aid
medicine.medical_specialty
Data collection
business.industry
Computer science
Hearing Loss, Sensorineural
Loudness Perception
medicine.medical_treatment
Repeated measures design
Audiogram
Audiology
Speech and Hearing
Hearing Aids
Software
Prosthesis Fitting
Prohibitins
Speech Perception
medicine
Humans
Research article
business
Insertion gain
Subjects
Details
- ISSN :
- 21573107 and 10500545
- Volume :
- 32
- Database :
- OpenAIRE
- Journal :
- Journal of the American Academy of Audiology
- Accession number :
- edsair.doi.dedup.....adda3b7aebc802b5603cad344deb3eb2
- Full Text :
- https://doi.org/10.1055/s-0041-1722947