Back to Search
Start Over
Development of an x-ray imaging system for the Laser Megajoule (LMJ)
- Source :
- Review of Scientific Instruments, Review of Scientific Instruments, 2010, 81 (10), pp.10E509. ⟨10.1063/1.3475788⟩, Review of Scientific Instruments, American Institute of Physics, 2010, 81 (10), pp.10E509. ⟨10.1063/1.3475788⟩
- Publication Year :
- 2010
- Publisher :
- HAL CCSD, 2010.
-
Abstract
- This imaging system aims at recording images of the core size and shape of an imploding deuterium-tritium (DT) microballoon on LMJ inertial confinement fusion (ICF) experiments. Image acquisition is difficult due to the harsh surrounding created by the fusion reaction, which affects system specifications. This one is made of a scintillator, an optical relay, and a CCD camera shielded from the surrounding. The system was tested on different facilities at CEA/DIF, where a spatial resolution of 120 μm was achieved and gamma dose up to 20 rad effects were measured. Setup and performed test are described.
- Subjects :
- 010302 applied physics
Physics
business.industry
02 engineering and technology
Scintillator
021001 nanoscience & nanotechnology
Laser
01 natural sciences
7. Clean energy
law.invention
Optics
law
0103 physical sciences
Shielded cable
[PHYS.HEXP]Physics [physics]/High Energy Physics - Experiment [hep-ex]
Nuclear fusion
Plasma diagnostics
0210 nano-technology
business
Instrumentation
Image resolution
Inertial confinement fusion
ComputingMilieux_MISCELLANEOUS
Laser Mégajoule
Subjects
Details
- Language :
- English
- ISSN :
- 00346748 and 10897623
- Database :
- OpenAIRE
- Journal :
- Review of Scientific Instruments, Review of Scientific Instruments, 2010, 81 (10), pp.10E509. ⟨10.1063/1.3475788⟩, Review of Scientific Instruments, American Institute of Physics, 2010, 81 (10), pp.10E509. ⟨10.1063/1.3475788⟩
- Accession number :
- edsair.doi.dedup.....ada3c404a864e3312d07a5dec7eae404
- Full Text :
- https://doi.org/10.1063/1.3475788⟩