- Høyer, C, Sandermann, J, Paludan, J P D, Pavar, S & Petersen, L J 2013, ' Diagnostic accuracy of laser Doppler flowmetry versus strain gauge plethysmography for segmental pressure measurement ', Journal of Vascular Surgery, vol. 58, no. 6, pp. 1563-70 . https://doi.org/10.1016/j.jvs.2013.06.057, Høyer, C, Sandermann, J, Paludan, J P D, Pavar, S & Petersen, L J 2013, ' Diagnostic accuracy of laser Doppler flowmetry versus strain gauge plethysmography for segmental pressure measurement ', Journal of Vascular Surgery, vol. 58, no. 6, pp. 1563-1570 . https://doi.org/10.1016/j.jvs.2013.06.057
- Publisher :
- Society for Vascular Surgery. Published by Elsevier Inc.
-
Abstract
- ObjectiveTo assess the diagnostic accuracy of laser Doppler flowmetry (LDF) with mercury-in-silastic strain gauge plethysmography (SGP) as a reference test for measuring the toe and ankle pressures in patients with known or suspected peripheral arterial disease (PAD).MethodsThis was a prospective, randomized, blinded diagnostic accuracy study. Toe and ankle pressures were measured using both methods in 200 consecutive patients, who were recruited at our vascular laboratory over a period of 30 working days. Classification of PAD and critical limb ischemia (CLI) was made in accordance with TASC-II criteria.ResultsThe LDF method demonstrated 5.8 mm Hg higher mean toe pressures than the SGP method for the right limb and 7.0 mm Hg for the left limb (both P < .001). There were no significant differences in the mean ankle pressures (both P > .129). The limits of agreement for the differences (SGP − LDF) were −31.7 to 20.2 mm Hg for right toe pressures, −28.0 to 14.0 mm Hg for left toe pressures, −25.5 to 22.8 mm Hg for right ankle pressures, and −26.9 to 24.6 mm Hg for left ankle pressures. A correlation analysis of the absolute pressures using the two methods showed an intraclass correlation coefficient of 0.902 (95% confidence interval [CI], 0.835-0.938) for right toe pressures, 0.919 (95% CI, 0.782-0.960) for the left toe pressures, 0.953 (95% CI, 0.937-0.965) for right ankle pressures, and 0.952 (95% CI, 0.936-0.964) for left ankle pressures. Cohen's Kappa showed an agreement in the diagnostic classification of κ = 0.775 (95% CI, 0.631-0.919) for PAD and κ = 0.780 (95% CI, 0.624-0.936) for CLI.ConclusionsLDF showed a good correlation with SGP over a wide range of toe and ankle pressures, as well as substantial agreement for the diagnostic classification of PAD including CLI.
- Subjects :
- Male
medicine.medical_specialty
Intraclass correlation
Diagnostic accuracy
Severity of Illness Index
Peripheral Arterial Disease
Double-Blind Method
medicine
Laser-Doppler Flowmetry
Plethysmograph
Humans
Plethysmography, Impedance
Prospective Studies
Aged
Leg
business.industry
Reproducibility of Results
Critical limb ischemia
Laser Doppler velocimetry
Prognosis
Confidence interval
Surgery
body regions
medicine.anatomical_structure
Female
Ankle
medicine.symptom
Cardiology and Cardiovascular Medicine
business
Nuclear medicine
Strain gauge plethysmography
Follow-Up Studies
Subjects
Details
- Language :
- English
- ISSN :
- 07415214
- Issue :
- 6
- Database :
- OpenAIRE
- Journal :
- Journal of Vascular Surgery
- Accession number :
- edsair.doi.dedup.....ad73414df1f1c3bb22d9c5fef1dcb4c4
- Full Text :
- https://doi.org/10.1016/j.jvs.2013.06.057