- The study analysed the mechanical properties by uniaxial test of porcine and bovine pericardial specimens, stabilised with the glutaraldehyde or with the ethylene glycol diglycidyl ether. Also the paper describes, the ability of these samples compressed to 18 Fr with via the finite element method. Analysis of the properties of the biomaterial demonstrated a low strength for all types of porcine pericardium with the lowest value for the EGDE-treated pericardium (2.75 N). The lowest stress-strain state during the simulation of the leaflet crimping was observed in experimental porcine pericardial patch with the minimum principal logarithmic strain of 0.32 m/m. The results of this study allow us to make conclusions about the desirability of the use of porcine pericardium treated with glutaraldehyde for transcatheter aortic valve prostheses.
- Subjects :
- medicine.medical_specialty
Pericardial patch
Materials science
Transcatheter aortic
Biomedical Engineering
Biomaterial
02 engineering and technology
021001 nanoscience & nanotechnology
Surgery
03 medical and health sciences
chemistry.chemical_compound
0302 clinical medicine
medicine.anatomical_structure
chemistry
cardiovascular system
medicine
Pericardium
Glutaraldehyde
0210 nano-technology
030217 neurology & neurosurgery
Biomedical engineering
Ethylene glycol diglycidyl ether
Fixation (histology)
Subjects
Details
- ISSN :
- 17526426 and 17526418
- Volume :
- 25
- Database :
- OpenAIRE
- Journal :
- International Journal of Biomedical Engineering and Technology
- Accession number :
- edsair.doi.dedup.....ac72582a1b90263321f82e73ccb879e4