Assessment of cardiac implantable electric device lead perforation using a metal artifact reduction algorithm in cardiac computed tomography

Authors :
Toshinori Hirai
Minako Azuma
Yasunori Nagayama
Hisanori Kanazawa
Takeshi Nakaura
Seitaro Oda
Seiji Takashio
Akira Sasao
Masafumi Kidoh
Koichi Kaikita
Kenichi Tsujita
Daisuke Sakabe
Masahiro Hatemura
Osamu Ikeda
Kengo Nakato
Yoshinori Funama
Source :
European Journal of Radiology. 136:109530
Publication Year :
2021
Publisher :
Elsevier BV, 2021.

Abstract

CT is considered the non-invasive gold standard for evaluating cardiac implantable electronic devices (CIEDs) lead perforation, but metal artifacts caused by the lead tip affect the image quality and make a definitive diagnosis challenging. We compared the performances of the metal artifact reduction (MAR) algorithm and the conventional algorithm for identification of the right ventricular (RV) lead tip position in cardiac CT studies of patients with CIEDs.Forty-seven consecutive patients (26 men; age 70.3 ± 15.4 years) with CIEDs underwent cardiac CT. Using the conventional and MAR algorithm, two image reconstructions were performed for each scan. We calculated the artifact index (AI) to assess the quantitative capability of the MAR algorithm for artifact reduction and visually assessed the RV lead tip position on both images as follows: non-perforation, perforation, and equivocal.The mean AIs were significantly lower with the MAR algorithm than with the conventional algorithm (96.7 ± 40.1 HU vs. 284.6 ± 134.1 HU, P 0.001). Thirteen (27.7 %) patients were diagnosed as equivocal using the conventional algorithm but were diagnosed with perforation (2 patients) and non-perforation (11 patients) using the MAR algorithm (equivocal rate: 27.7 % vs. 0%, P 0.001). Using the MAR algorithm, all cases were diagnosed with perforation (6 patients, 12.8 %) or non-perforation (41 patients, 87.2 %).The MAR algorithm effectively reduced metal artifacts and allowed us to diagnose the presence or absence of perforation in all cases, whereas definitive diagnosis was difficult with the use of conventional algorithm in 27.7 % of cases.

Details

ISSN :
0720048X
Volume :
136
Database :
OpenAIRE
Journal :
European Journal of Radiology
Accession number :
edsair.doi.dedup.....aa809002252f24a46fbd9aca1e8a2c1c