Back to Search Start Over

Measurement of the deuterium Balmer series line emission on EAST

Authors :
Wu Zebin
East Team
C. R. Wu
Peng Zhang
J. F. Chang
Yuqian Chen
Z. Xu
Y. M. Duan
Wen Gao
Zhan Jin
J. C. Xu
Y. M. Hou
Jiuyuan Li
Junchao Huang
Lingxuan Zhang
Source :
The Review of scientific instruments. 87(11)
Publication Year :
2016

Abstract

Volume recombination plays an important role towards plasma detachment for magnetically confined fusion devices. High quantum number states of the Balmer series of deuterium are used to study recombination. On EAST (Experimental Advanced Superconducting Tokamak), two visible spectroscopic measurements are applied for the upper/lower divertor with 13 channels, respectively. Both systems are coupled with Princeton Instruments ProEM EMCCD 1024B camera: one is equipped on an Acton SP2750 spectrometer, which has a high spectral resolution ∼0.0049 nm with 2400 gr/mm grating to measure the Dα(Hα) spectral line and with 1200 gr/mm grating to measure deuterium molecular Fulcher band emissions and another is equipped on IsoPlane SCT320 using 600 gr/mm to measure high-n Balmer series emission lines, allowing us to study volume recombination on EAST and to obtain the related line averaged plasma parameters (Te, ne) during EAST detached phases. This paper will present the details of the measurements and the characteristics of deuterium Balmer series line emissions during density ramp-up L-mode USN plasma on EAST.

Details

ISSN :
10897623
Volume :
87
Issue :
11
Database :
OpenAIRE
Journal :
The Review of scientific instruments
Accession number :
edsair.doi.dedup.....a9d75dc834b19730f35b163da5bb5dcc