Back to Search
Start Over
Preclinical assessment of a modified Occlutech left atrial appendage closure device in a canine model
- Source :
- International Journal of Cardiology. 221:413-418
- Publication Year :
- 2016
- Publisher :
- Elsevier BV, 2016.
-
Abstract
- Background LAA occlusion has a similar stroke prevention efficacy compared to anticoagulation treatment for non-valvular atrial fibrillation. Objective The objective of this study was to assess the feasibility and safety of a modified Occlutech® left atrial appendage (LAA) closure device in a canine model. Methods The device was implanted in 10 dogs (33±1kg) using fluoroscopy and transesophageal echocardiography (TEE) guidance. The modified Occlutech® LAA occlusion device was compared with the current version, the Watchman device, and the Amplazter cardiac plug (ACP). LAA occlusion and anchoring to the LAA were evaluated. All dogs were assessed using angiography, TEE, and a gross anatomy examination. Results The 10 LAA occlusion devices were to be implanted into 10 dogs (5 modified Occlutech devices, 3 current version of Occlutech devices, 1 Watchman, and 1 ACP). LAA implantation was not performed in one dog due to transeptal puncture failure. The three current version of Occlutech devices were embolized immediately after implantation, so three modified devices of the same size were implanted securely without embolization. The mean implant size was 20.1±2.0mm. The devices chosen were a mean of 23.3±10.6% larger than the measured landing zone diameters. Post-implant angiography and TEE revealed well-positioned devices without pericardial effusion or impingement on surrounding structures. Conclusions The results of this acute animal study suggested that a modified Occlutech® LAA occlusion device was feasible and had greater anchoring performance in canines. Additional large clinical studies are needed to evaluate safety and efficacy.
- Subjects :
- medicine.medical_specialty
Septal Occluder Device
medicine.medical_treatment
030204 cardiovascular system & hematology
Pericardial effusion
Prosthesis Implantation
03 medical and health sciences
Dogs
0302 clinical medicine
Left atrial
Internal medicine
Atrial Fibrillation
Occlusion
medicine
Animals
Fluoroscopy
Atrial Appendage
030212 general & internal medicine
Embolization
medicine.diagnostic_test
business.industry
Atrial fibrillation
medicine.disease
Stroke
Disease Models, Animal
Angiography
Cardiology
Cardiology and Cardiovascular Medicine
business
Canine model
Subjects
Details
- ISSN :
- 01675273
- Volume :
- 221
- Database :
- OpenAIRE
- Journal :
- International Journal of Cardiology
- Accession number :
- edsair.doi.dedup.....a99176cda63a8b541767619709e57302
- Full Text :
- https://doi.org/10.1016/j.ijcard.2016.07.048