Back to Search Start Over

RF Heating of Gold Cup and Conductive Plastic Electrodes during Simultaneous EEG and MRI

Authors :
Patrick Britz
John R. Ives
William M. Wells
Mukund Balasubramanian
Darren B. Orbach
Robert V. Mulkern
Source :
The Neurodiagnostic Journal. 57:69-83
Publication Year :
2017
Publisher :
Informa UK Limited, 2017.

Abstract

To investigate the heating of EEG electrodes during magnetic resonance imaging (MRI) scans and to better understand the underlying physical mechanisms with a focus on the antenna effect.Gold cup and conductive plastic electrodes were placed on small watermelons with fiberoptic probes used to measure electrode temperature changes during a variety of 1.5T and 3T MRI scans. A subset of these experiments was repeated on a healthy human volunteer.The differences between gold and plastic electrodes did not appear to be practically significant. For both electrode types, we observed heating below 4°C for straight wires whose lengths were multiples of ½ the radiofrequency (RF) wavelength and stronger heating (over 15°C) for wire lengths that were odd multiples of ¼ RF wavelength, consistent with the antenna effect.The antenna effect, which has received little attention so far in the context of EEG-MRI safety, can play as significant a role as the loop effect (from electromagnetic induction) in the heating of EEG electrodes, and therefore wire lengths that are odd multiples of ¼ RF wavelength should be avoided. These results have important implications for the design of EEG electrodes and MRI studies as they help to minimize the risk to patients undergoing MRI with EEG electrodes in place.

Details

ISSN :
23758627 and 21646821
Volume :
57
Database :
OpenAIRE
Journal :
The Neurodiagnostic Journal
Accession number :
edsair.doi.dedup.....a87d7ab9542e6f9443ce2d05f29850ba
Full Text :
https://doi.org/10.1080/21646821.2017.1256722