Back to Search Start Over

In-flight calibration and verification of the Planck-LFI instrument

Authors :
M. Frailis
R. C. Butler
P. Battaglia
Michele Maris
Christopher J. Watson
Luis Mendes
P. Wilson
Rodrigo Leonardi
M. Tomasi
Mike Krassenburg
Marco Bersanelli
Gianluca Morgante
Steve Foley
Frederick Melot
Anna Gregorio
J. A. Tauber
A. Mennella
Andrea Zonca
Daniele Tavagnacco
Althea Wilkinson
Andrea Zacchei
Dave Pearson
Damien Texier
Leticia Perez Cuevas
Maryse Charra
Thierry Maciaszek
Peter Meinhold
Richard J. Davis
Mauro Casale
Luca Terenzi
E. Franceschi
Michelle Baker
Steve Lowe
René J. Laureijs
Patrick Rihet
S. Galeotta
Luca Stringhetti
Fabrizio Villa
Patrick Stassi
Cristian Franceschet
F. Cuttaia
M. Sandri
Nazzareno Mandolesi
Laboratoire de Physique Subatomique et de Cosmologie (LPSC)
Institut polytechnique de Grenoble - Grenoble Institute of Technology (Grenoble INP )-Institut Polytechnique de Grenoble - Grenoble Institute of Technology-Institut National de Physique Nucléaire et de Physique des Particules du CNRS (IN2P3)-Université Joseph Fourier - Grenoble 1 (UJF)-Centre National de la Recherche Scientifique (CNRS)
Institut d'astrophysique spatiale (IAS)
Université Paris-Sud - Paris 11 (UP11)-Institut national des sciences de l'Univers (INSU - CNRS)-Centre National de la Recherche Scientifique (CNRS)
Centre National d'Études Spatiales [Toulouse] (CNES)
PLANCK
Centre National de la Recherche Scientifique (CNRS)-Institut national des sciences de l'Univers (INSU - CNRS)-Université Paris-Sud - Paris 11 (UP11)
Gregorio, Anna
F., Cuttaia
A., Mennella
M., Bersanelli
M., Mari
P., Meinhold
M., Sandri
L., Terenzi
M., Tomasi
F., Villa
M., Fraili
G., Morgante
D., Pearson
A., Zacchei
P., Battaglia
R. C., Butler
R., Davi
C., Franceschet
E., Franceschi
S., Galeotta
R., Leonardi
S., Lowe
N., Mandolesi
F., Melot
L., Mende
P., Stassi
L., Stringhetti
Tavagnacco, Daniele
A., Zonca
A., Wilkinson
P., Wilson
M., Charra
T., Maciaszek
S., Foley
C. J., Watson
M., Casale
R., Laureij
J., Tauber
D., Texier
M., Baker
L., Perez Cueva
M., Krassenburg
P., Rihet
Source :
Journal of Instrumentation, Journal of Instrumentation, IOP Publishing, 2013, 8, pp.T07001. ⟨10.1088/1748-0221/8/07/T07001⟩
Publication Year :
2013
Publisher :
arXiv, 2013.

Abstract

In this paper we discuss the Planck-LFI in-flight calibration campaign. After a brief overview of the ground test campaigns, we describe in detail the calibration and performance verification (CPV) phase, carried out in space during and just after the cool-down of LFI. We discuss in detail the functionality verification, the tuning of the front-end and warm electronics, the preliminary performance assessment and the thermal susceptibility tests. The logic, sequence, goals and results of the in-flight tests are discussed. All the calibration activities were successfully carried out and the instrument response was comparable to the one observed on ground. For some channels the in-flight tuning activity allowed us to improve significantly the noise performance.<br />Comment: Long technical paper on Planck LFI in flight calibration campaign: 109 pages in this (not final) version, 100 page in the final JINST version

Details

ISSN :
17480221
Database :
OpenAIRE
Journal :
Journal of Instrumentation, Journal of Instrumentation, IOP Publishing, 2013, 8, pp.T07001. ⟨10.1088/1748-0221/8/07/T07001⟩
Accession number :
edsair.doi.dedup.....a832689b225b21e6247d1ed075c8db41
Full Text :
https://doi.org/10.48550/arxiv.1307.2029