An innovative reactor-type biosensor for BOD rapid measurement
- Source :
- Biosensorsbioelectronics. 25(7)
- Publication Year :
- 2009
-
Abstract
- Biochemical oxygen demand (BOD) is one of the most important and widely used parameters for characterizing the organic pollution of water and wastewater. In this paper, a novel reactor-type biosensor for rapid measurement of BOD was developed, based on using immobilized microbial cell (IMC) beads as recognition bio-element in a completely mixed reactor which was used as determining chamber, replacing the traditionally used membrane as recognition bio-element. The IMC beads were freely suspended in the aqueous solution, so the mass transfer resistance for dissolved oxygen and organic compounds significantly reduced, and the quantity of the microbial cells used as recognition element can be easily adjusted, in comparison with the traditional membrane-type BOD biosensor, in which exists a unadjustable contradiction between the quantity of biomass and the thickness of the bio-membrane, thus limiting the stability and the detection limit. This novel kind of BOD biosensor significantly increased the sensitivity of the response, the detecting precision and prolonged the life time of the recognition element. The experimental data showed that the most appropriate temperature for biochemical reaction in the reactor was 30 degrees C, and the IMC beads could keep the bioactivity for about 70d at the detecting frequency of 8 times every day. The standard deviation of repeatability and the reproducibility of responses were within +/-6.4% and +/-5.0%, respectively, which are within acceptable bias limits, and meet the requirement of BOD rapid measurement.
- Subjects :
- Biochemical oxygen demand
Biomedical Engineering
Biophysics
Analytical chemistry
Biosensing Techniques
Bacterial Physiological Phenomena
Sensitivity and Specificity
Bioreactors
Mass transfer
Electrochemistry
Water Pollutants
Oximetry
Detection limit
Reproducibility
Aqueous solution
Chromatography
Chemistry
Reproducibility of Results
General Medicine
Repeatability
Equipment Design
Equipment Failure Analysis
Oxygen
Wastewater
Biological Assay
Water Microbiology
Biosensor
Biotechnology
Subjects
Details
- ISSN :
- 18734235
- Volume :
- 25
- Issue :
- 7
- Database :
- OpenAIRE
- Journal :
- Biosensorsbioelectronics
- Accession number :
- edsair.doi.dedup.....a6e018779bdc2de0b28042b932b3c5a2