Experimental Validation of a Gamma Detector With a Novel Light-Guide-PMT Geometry to Reduce Dead Edge Effects
- Source :
- IEEE Transactions on Radiation and Plasma Medical Sciences, 4(1)
- Publication Year :
- 2020
- Publisher :
- Institute of Electrical and Electronics Engineers (IEEE), 2020.
-
Abstract
- Photomultiplier tube (PMT)-based scintillation cameras are predominant in molecular imaging but have the drawback that position estimation is severely degraded near the edges (dead edge effect). This leads to sensitivity losses and can cause severe problems in applications like molecular breast imaging and in certain SPECT devices. Using smaller light sensors or semiconductor detectors can solve this issue but leads to increased costs. Here we present a gamma detector based on standard PMTs with a novel light-guide-PMT geometry that strongly reduces dead edges. In our design, a monolithic NaI(Tl) scintillator is read out by square PMTs placed in a staggered arrangement. At the edge of the scintillator we inserted additional light-guides to emulate half-size PMTs. Detector performance was assessed for 99m Tc imaging; an average spatial resolution of 3.6 mm was measured over the whole detector, degrading to 4.0 mm within 30 mm to the critical edge. The dead edge of the scintillator is
- Subjects :
- Physics
Photomultiplier
Anger camera
Detector
dead edge
Geometry
Scintillator
Edge (geometry)
Atomic and Molecular Physics, and Optics
light-guide
Anger Camera
Semiconductor detector
scintillator
molecular breast imaging
Radiology, Nuclear Medicine and imaging
gamma detector
Instrumentation
Sensitivity (electronics)
Image resolution
Subjects
Details
- ISSN :
- 24697303 and 24697311
- Volume :
- 4
- Database :
- OpenAIRE
- Journal :
- IEEE Transactions on Radiation and Plasma Medical Sciences
- Accession number :
- edsair.doi.dedup.....a6453a94205516879d20d80893691a65
- Full Text :
- https://doi.org/10.1109/trpms.2019.2916386