Accuracy and precision of four value-added blood glucose meters: the Abbott Optium, the DDI Prodigy, the HDI True Track, and the HypoGuard Assure Pro
- Source :
- Diabetes technologytherapeutics. 11(9)
- Publication Year :
- 2009
-
Abstract
- This study compared the accuracy and precision of four value-added glucose meters.Finger stick glucose measurements in diabetes patients were performed using the Abbott Diabetes Care (Alameda, CA) Optium, Diagnostic Devices, Inc. (Miami, FL) DDI Prodigy, Home Diagnostics, Inc. (Fort Lauderdale, FL) HDI True Track Smart System, and Arkray, USA (Minneapolis, MN) HypoGuard Assure Pro. Finger glucose measurements were compared with laboratory reference results. Accuracy was assessed by a Clarke error grid analysis (EGA), a Parkes EGA, and within 5%, 10%, 15%, and 20% of the laboratory value criteria (chi2 analysis). Meter precision was determined by calculating absolute mean differences in glucose values between duplicate samples (Kruskal-Wallis test).Finger sticks were obtained from 125 diabetes patients, of which 90.4% were Caucasian, 51.2% were female, 83.2% had type 2 diabetes, and average age of 59 years (SD 14 years). Mean venipuncture blood glucose was 151 mg/dL (SD +/-65 mg/dL; range, 58-474 mg/dL). Clinical accuracy by Clarke EGA was demonstrated in 94% of Optium, 82% of Prodigy, 61% of True Track, and 77% of the Assure Pro samples (P0.05 for Optium and True Track compared to all others). By Parkes EGA, the True Track was significantly less accurate than the other meters. Within 5% accuracy was achieved in 34%, 24%, 29%, and 13%, respectively (P0.05 for Optium, Prodigy, and Assure Pro compared to True Track). Within 10% accuracy was significantly greater for the Optium, Prodigy, and Assure Pro compared to True Track. Significantly more Optium results demonstrated within 15% and 20% accuracy compared to the other meter systems. The HDI True Track was significantly less precise than the other meter systems.The Abbott Optium was significantly more accurate than the other meter systems, whereas the HDI True Track was significantly less accurate and less precise compared to the other meter systems.
- Subjects :
- Gerontology
Blood Glucose
Male
Accuracy and precision
Time Factors
Endocrinology, Diabetes and Metabolism
Grid analysis
Type 2 diabetes
Statistics, Nonparametric
Finger Stick
Endocrinology
Diabetes mellitus
medicine
Confidence Intervals
Diabetes Mellitus
Humans
Insulin
Aged
Blood glucose meters
Analysis of Variance
Blood Specimen Collection
Venipuncture
Chi-Square Distribution
business.industry
Blood Glucose Self-Monitoring
Glucose Measurement
Middle Aged
medicine.disease
United States
Diet
Medical Laboratory Technology
Female
business
Nuclear medicine
Subjects
Details
- ISSN :
- 15578593
- Volume :
- 11
- Issue :
- 9
- Database :
- OpenAIRE
- Journal :
- Diabetes technologytherapeutics
- Accession number :
- edsair.doi.dedup.....a5715a32955be39f222cc94a12c4cb52