Accurate 3D temperature dosimetry during hyperthermia therapy by combining invasive measurements and patient-specific simulations

Authors :
Gerda M. Verduijn
Z. Rijnen
Margarethus M. Paulides
Jifke F. Veenland
Valerio Fortunati
Theo van Walsum
René F. Verhaart
Biomedical Engineering
Radiotherapy
Radiology & Nuclear Medicine
Medical Informatics
Source :
International Journal of Hyperthermia, 31(6), 686-692. Informa Healthcare
Publication Year :
2015

Abstract

Purpose: Dosimetry during deep local hyperthermia treatments in the head and neck currently relies on a limited number of invasively placed temperature sensors. The purpose of this study was to assess the feasibility of 3D dosimetry based on patient-specific temperature simulations and sensory feedback. Materials and methods: The study includes 10 patients with invasive thermometry applied in at least two treatments. Based on their invasive thermometry, we optimised patient-group thermal conductivity and perfusion values for muscle, fat and tumour using a 'leave-one-out' approach. Next, we compared the accuracy of the predicted temperature (Delta T) and the hyperthermia treatment quality (Delta T50) of the optimisations based on the patient-group properties to those based on patient-specific properties, which were optimised using previous treatment measurements. As a robustness check, and to enable comparisons with previous studies, we optimised the parameters not only for an applicator efficiency factor of 40%, but also for 100% efficiency. Results: The accuracy of the predicted temperature (Delta T) improved significantly using patient-specific tissue properties, i.e. 1.0 degrees C (inter-quartile range (IQR) 0.8 degrees C) compared to 1.3 degrees C (IQR 0.7 degrees C) for patient-group averaged tissue properties for 100% applicator efficiency. A similar accuracy was found for optimisations using an applicator efficiency factor of 40%, indicating the robustness of the optimisation method. Moreover, in eight patients with repeated measurements in the target region, Delta T50 significantly improved, i.e. Delta T50 reduced from 0.9 degrees C (IQR 0.8 degrees C) to 0.4 degrees C (IQR 0.5 degrees C) using an applicator efficiency factor of 40%. Conclusion: This study shows that patient-specific temperature simulations combined with tissue property reconstruction from sensory data provides accurate minimally invasive 3D dosimetry during hyperthermia treatments: T50 in sessions without invasive measurements can be predicted with a median accuracy of 0.4 degrees C.

Details

ISSN :
14645157 and 02656736
Volume :
31
Issue :
6
Database :
OpenAIRE
Journal :
International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group
Accession number :