Back to Search Start Over

Combined use of HPLC-ICP-MS and microwave-assisted extraction for the determination of cobalt compounds in nutritive supplements

Authors :
A.C. Sahayam
Shiuh-Jen Jiang
Fang-Yu Yang
Source :
Food chemistry. 147
Publication Year :
2012

Abstract

Speciation analysis of cobalt in nutritive supplements has been carried out using HPLC and ICP-MS equipped with a membrane desolvation sample introduction system as detector. In this study, cobalt containing compounds, namely Co(II), cyanocobalamin (CN-Cbl) and hydroxylcobalamin (OH-Cbl), were well separated by reversed phase HPLC with a C8-HPLC column as the stationary phase and 8 mmol L(-1) ammonium acetate in 22%v/v methanol solution (pH 4) as the mobile phase using isocratic elution. Detection limit was in the range of 0.008-0.014 μg CoL(-1) for various Co species. Over 98% of the total cobalt species was extracted in nutritive supplements using a 0.5%v/v HNO3 solution in a microwave field; and the spike recovery was in the range of 92-108% for various species. The HPLC-ICP-MS results showed a satisfactory agreement with the total cobalt concentrations obtained by ICP-MS analysis of completely dissolved samples.

Details

ISSN :
18737072
Volume :
147
Database :
OpenAIRE
Journal :
Food chemistry
Accession number :
edsair.doi.dedup.....a46ec82d1f1239eb232b21ff86b1cf0f