A Dosimetric evaluation of a high dose rate cobalt-60 brachytherapy source using shielded, single and multi-channel cylinder applicators for gynecological cancers
- Source :
- Medical Dosimetry. 47:318-324
- Publication Year :
- 2022
- Publisher :
- Elsevier BV, 2022.
-
Abstract
- Vaginal brachytherapy is a standard method for preventing the recurrences of malignancies in the treatment of gynecological cancers. Cylindrical applicators are usually employed in high dose rate cobalt-60 source brachytherapy after abdominal hysterectomy or bilateral salpingo-oophorectomy treatment. The aim of this study is to conduct an assessment of dosimetric properties of single channel, multichannel, and shielded cylinder applicators. This study utilizes a pelvis phantom with three different volumes of bladder and rectum to perform dose measurement around the cylindrical applicators. GAFCHROMIC EBT3 model film was used to measure dose distributions in water phantom around the bladder, rectum, applicator, and CTV. In order to properly compare the performance of the shielded applicator with multichannel and single channel applicators, we consider the both cases of a one-sided CTV and two CTVs placed on both sides of the vagina. Dose-volume histograms analysis shows no statistically significant difference in tumoral region for the two types of CTV between cylindrical applicators. The V