Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study
- Source :
- Radiology. 287:224-234
- Publication Year :
- 2018
- Publisher :
- Radiological Society of North America (RSNA), 2018.
-
Abstract
- Purpose To determine the accuracy of dual-energy computed tomographic (CT) quantitation in a phantom system comparing fast kilovolt peak-switching, dual-source, split-filter, sequential-scanning, and dual-layer detector systems. Materials and Methods A large elliptical phantom containing iodine (2, 5, and 15 mg/mL), simulated contrast material-enhanced blood, and soft-tissue inserts with known elemental compositions was scanned three to five times with seven dual-energy CT systems and a total of 10 kilovolt peak settings. Monochromatic images (50, 70, and 140 keV) and iodine concentration images were created. Mean iodine concentration and monochromatic attenuation for each insert and reconstruction energy level were recorded. Measurement bias was assessed by using the sum of the mean signed errors measured across relevant inserts for each monochromatic energy level and iodine concentration. Iodine and monochromatic errors were assessed by using the root sum of the squared error of all measurements. Results At least one acquisition paradigm per scanner had iodine biases (range, -2.6 to 1.5 mg/mL) with significant differences from zero. There were no significant differences in iodine error (range, 0.44-1.70 mg/mL) among the top five acquisition paradigms (one fast kilovolt peak switching, three dual source, and one sequential scanning). Monochromatic bias was smallest for 70 keV (-12.7 to 15.8 HU) and largest for 50 keV (-80.6 to 35.2 HU). There were no significant differences in monochromatic error (range, 11.4-52.0 HU) among the top three acquisition paradigms (one dual source and two fast kilovolt peak switching). The lowest accuracy for both measures was with a split-filter system. Conclusion Iodine and monochromatic accuracy varies among systems, but dual-source and fast kilovolt-switching generally provided the most accurate results in a large phantom. © RSNA, 2017 Online supplemental material is available for this article.
- Subjects :
- Scanner
Mean squared error
Phantoms, Imaging
business.industry
Attenuation
Detector
Reproducibility of Results
Imaging phantom
030218 nuclear medicine & medical imaging
Radiographic Image Enhancement
Radiography, Dual-Energy Scanned Projection
03 medical and health sciences
0302 clinical medicine
Optics
030220 oncology & carcinogenesis
Radiographic Image Interpretation, Computer-Assisted
Medicine
Radiology, Nuclear Medicine and imaging
Tomography
Monochromatic color
Tomography, X-Ray Computed
business
Nuclear medicine
Iodine
Kilovolt Peak
Subjects
Details
- ISSN :
- 15271315 and 00338419
- Volume :
- 287
- Database :
- OpenAIRE
- Journal :
- Radiology
- Accession number :
- edsair.doi.dedup.....a3765369851169d8fcd4f84cc0da1c00