Technical Features and Clinical Outcomes of Coronary Venous Left Ventricular Lead Removal and Reimplantation
- Source :
- Circulation journal : official journal of the Japanese Circulation Society. 85(8)
- Publication Year :
- 2021
-
Abstract
- Background The number of patients undergoing cardiac resynchronization therapy has increased. Consequently, there is increased frequency in the removal and reimplantation of coronary venous (CV) leads due to infection or malfunction.Methods and Results:A total of 345 consecutive patients referred for lead(s) extraction were reviewed. Of these, 34 patients who underwent a CV lead removal were investigated. The indications for CV leads removal were device-related infections in 29 patients and lead malfunctions in 5 patients. The average duration of the CV leads was 4.1±3.8 years. All CV leads were successfully removed without any major complications, except for 1 in-hospital death. Successful CV lead removal by simple traction (ST) was achieved in 21 patients (62%), whereas extraction tools were required in 13 patients (38%). Local infection and CV lead dwell time were significantly associated with successful ST (P=0.04 and P=0.014, respectively). CV lead re-implantation was successfully performed in 25 patients; however, a right-side approach was required in 92%, and occlusion/stenosis of the previous CV was observed in 80% of the patients. Conclusions CV lead removal is relatively successful and safe. The presence of local infection and a shorter lead duration may enable successful ST of a CV lead. However, the re-implantation procedure should be well prepared for the complexity related to the right-side approach and occlusion/stenosis of the previous CV.
- Subjects :
- medicine.medical_specialty
Average duration
Pacemaker, Artificial
Ventricular lead
medicine.medical_treatment
Cardiac resynchronization therapy
Constriction, Pathologic
030204 cardiovascular system & hematology
03 medical and health sciences
Local infection
0302 clinical medicine
Internal medicine
Occlusion
Medicine
Humans
030212 general & internal medicine
Major complication
Hospital Mortality
Lead (electronics)
Device Removal
business.industry
General Medicine
medicine.disease
Defibrillators, Implantable
Electrodes, Implanted
Stenosis
Treatment Outcome
Replantation
Cardiology
Cardiology and Cardiovascular Medicine
business
Subjects
Details
- ISSN :
- 13474820
- Volume :
- 85
- Issue :
- 8
- Database :
- OpenAIRE
- Journal :
- Circulation journal : official journal of the Japanese Circulation Society
- Accession number :
- edsair.doi.dedup.....a2a36789a04809aab62428ccb469f04c