Back to Search Start Over

Adaptive pulse oximeter with dual-wavelength based on wavelet transforms

Authors :
Shengjia Wang
Guangyu Li
Ziang Feng
Zhan Gao
Source :
Optics Express. 21:23058
Publication Year :
2013
Publisher :
The Optical Society, 2013.

Abstract

Pulse oximeter is widely used in the monitoring of blood oxygen in clinic for its convenience and efficiency. However, synchronizing light source flashing with data collecting is required, otherwise the separation of the data from different LEDs will fail. More importantly, synchronous acquisition makes the pulse oximetry system vulnerable. Meanwhile, the pulse waveform extraction is a crucial procedure in the measurement. Hence, in this paper, an asynchronous acquisition pulse oximetry system based on wavelet transform has been built. PhotoPlethysmoGraph (PPG) and photoelectric detection technology are applied in our homemade system. The adaptive soft-threshold de-noising is realized by Stein's Unbiased Risk Estimate (SURE). The principle and system configuration are described. The preliminary experiment results from wavelet transforms and Fourier transforms are compared. The results show that our homemade system is adaptive, accurate, robust and simple.

Details

ISSN :
10944087
Volume :
21
Database :
OpenAIRE
Journal :
Optics Express
Accession number :
edsair.doi.dedup.....9fd22302c0ffc07086baf78ac7d1e42e
Full Text :
https://doi.org/10.1364/oe.21.023058