Mechanical Characteristics of Dilated Polytetraflouroethylene Used for Transluminally Placed Endovascular Grafts
- Source :
- Annals of Vascular Surgery. 11:68-73
- Publication Year :
- 1997
- Publisher :
- Elsevier BV, 1997.
-
Abstract
- This study was conducted to assess the mechanical characteristics of dilated polytetraflouroethylene (PTFE) for use in transluminally placed endovascular grafts (TPEGs). Ten-centimeter lengths of 3- and 4-mm thinwalled PTFE were dilated to 8, 10, 12, and 15 mm diameters (3 mm) and 10-, 14-, 16-, and 20-mm diameters (4 mm), respectively (n = 6 for each size). The dilated PTFE segments were evaluated for leakage, further dilation, structural changes (with electron microscopy), and changes in wall thickness occurring after 24 hours of perfusion at pressures of 300-350 mmHg. Both 3- and 4-mm thinwalled PTFE could be dilated to five times their initial diameter before rupture occurred. Three-millimeter grafts dilated to 12- and 4-mm grafts dilated to 14 mm remained resistant to leakage at perfusion pressures up to 350 mmHg. When 3-mm grafts were dilated to 15 mm, the PTFE leaked saline at a rate of 20.3 +/- 9.3 cc per hour at 300 mmHg. pressure. Four-millimeter grafts dilated to 16- and 20-mm diameters leaked saline at 8.4 +/- 7.8 and 52.8 +/- 22 cc per minute, respectively, at the same pressure. No grafts were found to increase in diameter after 24 hours of pressure perfusion. Electron microscopy revealed that PTFE node size was significantly smaller in dilated grafts than in undilated grafts, but there was no significant change in internodal distance. This data suggests that thinwalled PTFE can be dilated to large diameters and retain sufficient strength to resist supraphysiologic pressures. Long-term studies are needed to determine the late structural integrity of dilated PTFE.
- Subjects :
- medicine.medical_specialty
Time Factors
medicine.medical_treatment
Catheterization
chemistry.chemical_compound
Blood vessel prosthesis
medicine
Humans
Polytetrafluoroethylene
Saline
business.industry
Structural integrity
General Medicine
Blood Vessel Prosthesis
Prosthesis Failure
Surgery
Perfusion
chemistry
Microscopy, Electron, Scanning
Balloon dilation
Stress, Mechanical
Cardiology and Cardiovascular Medicine
business
Nuclear medicine
Wall thickness
Subjects
Details
- ISSN :
- 08905096
- Volume :
- 11
- Database :
- OpenAIRE
- Journal :
- Annals of Vascular Surgery
- Accession number :
- edsair.doi.dedup.....9f2fe9be231a543933368917ae49060d
- Full Text :
- https://doi.org/10.1007/s100169900012