Back to Search Start Over

Point of care with micro fluidic paper based device integrated with nano zeolite–graphene oxide nanoflakes for electrochemical sensing of ketamine

Authors :
Dhritiman Chakraborty
Anusree Anil
Jagriti Narang
Ashish Mathur
Chandra Shekhar Pundir
Nitesh Malhotra
Aviraj Ingle
Chaitali Singhal
Source :
Biosensors and Bioelectronics. 88:249-257
Publication Year :
2017
Publisher :
Elsevier BV, 2017.

Abstract

The present study was aimed to develop an ultrasensitive technique for electroanalysis of ketamine; a date rape drug. It involved the fabrication of nano-hybrid based electrochemical micro fluidic paper-based analytical device (EμPADs) for electrochemical sensing of ketamine. A paper chip was developed using zeolites nanoflakes and graphene-oxide nanocrystals (Zeo–GO). EμPAD offers many advantages such as facile approach, economical and potential for commercialization. Nanocrystal modified EμPAD showed wide linear range 0.001–5 nM/mL and a very low detection limit of 0.001 nM/mL. The developed sensor was tested in real time samples like alcoholic and non-alcoholic drinks and found good correlation (99%). The hyphenation of EμPAD integrated with nanocrystalline Zeo–GO for detection of ketamine has immense prospective for field-testing platforms. An extensive development could be made for industrial translation of this fabricated device.