Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement
- Source :
- PLoS ONE, Vol 14, Iss 1, p e0210780 (2019), PLoS ONE
- Publication Year :
- 2019
- Publisher :
- Public Library of Science (PLoS), 2019.
-
Abstract
- The expanded polytetrafluoroethylene (ePTFE) heart valve can serve as a viable option for prosthetic aortic valve. In this study, an ePTFE bi-leaflet valve design for aortic valve replacement (AVR) is presented, and the performance of the proposed valve was assessed numerically and experimentally. The valve was designed using CAE software. The dynamic behavior of the newly designed bi-leaflet valve under time-varying physiological pressure loading was first investigated by using commercial finite element code. Then, in-vitro tests were performed to validate the simulation and to assess the hemodynamic performance of the proposed design. A tri-leaflet ePTFE valve was tested in-vitro under the same conditions as a reference. The maximum leaflet coaptation area of the bi-leaflet valve during diastole was 216.3 mm2. When fully closed, no leakage gap was observed and the free edges of the molded valve formed S-shaped lines. The maximum Von Mises stress during a full cardiac cycle was 4.20 MPa. The dynamic performance of the bi-leaflet valve was validated by the in-vitro test under physiological aortic pressure pulse. The effective orifice area (EOA), mean pressure gradient, regurgitant volume, leakage volume and energy loss of the proposed valve were 3.14 cm2, 8.74 mmHg, 5.93 ml/beat, 1.55 ml/beat and 98.99 mJ, respectively. This study reports a novel bi-leaflet valve design for AVR. The performance of the proposed valve was numerically and experimentally assessed. Compared with the reference valve, the proposed design exhibited better structural and hemodynamic performances, which improved valve competency. Moreover, the performance of the bi-leaflet design is comparable to commercialized valves available on the market. The results of the present study provide a viable option for the future clinical applications. NMRC (Natl Medical Research Council, S’pore) Published version
- Subjects :
- Aortic valve
030204 cardiovascular system & hematology
Pediatrics
chemistry.chemical_compound
Pediatric Surgery
0302 clinical medicine
Aortic valve replacement
Medicine and Health Sciences
Polytetrafluoroethylene
Aorta
Prosthetics
Multidisciplinary
Cardiac cycle
Physics
Models, Cardiovascular
Classical Mechanics
Heart
Hematology
Engineering::Mechanical engineering [DRNTU]
medicine.anatomical_structure
Heart Valve Prosthesis
Aortic Valve
Physical Sciences
cardiovascular system
Aortic pressure
Engineering and Technology
Mechanical Stress
Computer-Aided Design
Medicine
Anatomy
Research Article
Biotechnology
Materials science
Science
Finite Element Analysis
Diastole
Bioengineering
Surgical and Invasive Medical Procedures
In Vitro Techniques
Prosthesis Design
03 medical and health sciences
Coronary circulation
Imaging, Three-Dimensional
Coronary Circulation
medicine
Humans
Arterial Pressure
Computer Simulation
Heart valve
Hemodynamics
Biology and Life Sciences
medicine.disease
Assistive Technologies
030228 respiratory system
chemistry
Cardiovascular Anatomy
Blood Vessels
Medical Devices and Equipment
Stress, Mechanical
Biomedical engineering
Subjects
Details
- ISSN :
- 19326203
- Volume :
- 14
- Database :
- OpenAIRE
- Journal :
- PLOS ONE
- Accession number :
- edsair.doi.dedup.....9eb0650712f1a8878f8b590dc6230f88
- Full Text :
- https://doi.org/10.1371/journal.pone.0210780