Automation of sample mounting for macromolecular crystallography

Authors :
J.-S. Aksoy
M. Dallery
L. Launer
Julien Huet
F. di-Chiaro
M. Renier
Matias Guijarro
C. Taffut
J. E. McCarthy
Florent Cipriani
Gordon A. Leonard
Andrew Thompson
Raimond B. G. Ravelli
Sine Larsen
Stephen Cusack
Franck Felisaz
Hugo Caserotto
Mario Lentini
Martin A. Walsh
Sean McSweeney
Source :
Acta crystallographica. Section D, Biological crystallography. 62(Pt 10)
Publication Year :
2005

Abstract

A standard sample holder and vial for cryocooled macromolecular crystals has been defined for use with robotic sample changers. This SPINE standard sample holder is a modified version, with added features and specifications, of sample holders in common use. In particular, the SPINE standard meets the precision required for automatic sample exchange and includes a cap that is identified by a two-dimensional datamatrix code as well as an optional vial. At the ESRF, the sample holder standard is in use with the EMBL/ESRF/BM14 robotic sample changer (SC3) which is installed on eight beamlines. The SC3 can hold up to 50 crystals stored in five baskets. A datamatrix reader in the SC3 ensures safe management of the sample flow and facilitates fully automatic screening and characterization of samples. Tools for handling and transporting 50 samples in a dry shipping dewar have been developed. In addition to the SC3, the SPINE sample holder is currently compatible with a number of other robotic sample changers.

Details

ISSN :
09074449
Volume :
62
Issue :
Pt 10
Database :
OpenAIRE
Journal :
Acta crystallographica. Section D, Biological crystallography
Accession number :
edsair.doi.dedup.....9e93d7cf76f2412e783bafc7ce4c2046