Back to Search
Start Over
Strategies and analytical workflows to extend the dynamic range in quantitative LC-MS/MS analysis
- Source :
- Bioanalysis. 11(12)
- Publication Year :
- 2019
-
Abstract
- Aim: To evaluate alternative analytical strategies to extend the dynamic range in quantitative LC–MS/MS. Methods & results: Two approaches based on prior or no prior knowledge of expected exposure levels were evaluated. These approaches make use of two analytical strategies, which include the use of more than one injection volume or dilution of sample extract with solvents or solvent mixtures. A total of 16 compounds with varying logP values were classified into polar and nonpolar groups and used in this evaluation. From the two analytical strategies, three workflows were derived. Conclusion: All three workflows were successfully evaluated and resulted in good accuracy (80–120%) for all the compound groups.
- Subjects :
- Quality Control
Bioanalysis
Computer science
Dynamic range
010401 analytical chemistry
Clinical Biochemistry
Clinical Chemistry Tests
General Medicine
010402 general chemistry
01 natural sciences
0104 chemical sciences
Analytical Chemistry
Dilution
Workflow
Medical Laboratory Technology
Tandem Mass Spectrometry
Lc ms ms
Injection volume
Calibration
General Pharmacology, Toxicology and Pharmaceutics
Biological system
Chromatography, Liquid
Subjects
Details
- ISSN :
- 17576199
- Volume :
- 11
- Issue :
- 12
- Database :
- OpenAIRE
- Journal :
- Bioanalysis
- Accession number :
- edsair.doi.dedup.....9ddbf2d0db33b1ad3d1082cbbf2cb9ca