Comparison of a fluorometric assay kit with high-performance liquid chromatography for the assessment of serum retinol concentration

Authors :
El Haloui Noureddine
Azlaf Mehdi
El Hamdouchi Asmaa
Aguenaou Hassan
Benjeddou Kaoutar
Aglago Kouassivi Elom
El Menchawy Imane
El Kari Khalid
Source :
African Health Sciences; Vol 15, No 2 (2015); 641-646
Publication Year :
2015
Publisher :
Makerere University Medical School (Uganda), 2015.

Abstract

Background : Although high-performance liquid chromatography (HPLC) is the commonly used method for the analysis of retinol in biological samples, simple and rapid test kits are available. Objectives : This study compared a rapid test kit (ICHECK Fluoro®) to HPLC for the assessment of serum retinol concentrations. Methods : For the analysis by HPLC, sample preparation included standard deproteinization and extraction phases. The analysis by ICHECK was performed by injecting serum into IEX reagent vials (n=89) and mixing manually for separation. After precipitation of the proteins, the vial was introduced into the chamber of the ICHECK Fluoro and analysed at 0 min (ICHECK0min) and 15 min later (ICHECK15min). Bland and Altman approach was applied to test the agreement between HPLC and ICHECK. Results : Mean HPLC, ICHECK0min and ICHECK15min values were 421.2±106.0 μg/L, 423.1±118.3 μg/L and 413.2±107.6 μg/L, respectively. Retinol concentrations significantly decreased in the IEX solution over time (p