Agreement of white-to-white measurements with swept-source OCT, Scheimpflug and color LED devices

Authors :
Pedro Tañá-Rivero
Robert Montés-Micó
Salvador Aguilar-Córcoles
Ramón Ruiz-Mesa
Jose L. Rodriguez-Prats
Source :
International ophthalmology. 41(1)
Publication Year :
2020

Abstract

To assess the interchangeability of different devices for measuring white-to-white (WTW) distance. WTW distance was measured in 53 eyes of 53 patients using Anterion swept-source optical coherence topographer (SS-OCT), IOLMaster 700 SS-OCT, Pentacam HR Scheimpflug and Cassini color LED. Statistical analysis was done by means of the Friedman test and the post hoc Tukey test. The Bland–Altman analysis was applied to carry out pairwise comparisons with the average difference, 95% confidence interval of the average difference and limits of agreement 95% (LoA). WTW values obtained by the Anterion, IOLMaster 700, Pentacam HR and Cassini were: 11.84 ± 0.41 mm, 11.96 ± 0.41 mm, 11.68 ± 0.38 mm and 12.65 ± 0.52 mm, respectively. Statistically significant differences were found in all pairwise comparison (p