Back to Search
Start Over
Electrochemical ELISA Based on E. Coli with Autodisplayed Z-Domains
- Source :
- Procedia Engineering. 25:944-947
- Publication Year :
- 2011
- Publisher :
- Elsevier BV, 2011.
-
Abstract
- Recently, we reported autodisplayed Z-domains on E.coli can highly improve the sensitivity of immunoassays through the orientation control of detection antibodies [1–3]. In this work, the amperometric analysis was applied to the immunoassay based on E.coli cells with autodisplayed Z-domains. For amperometric analaysis, the reduction current of 3,5,3’,5’-tetramethylbenzidine (TMB) was measured at the potential of -50mV and the correlation between the reduction current and the OD value was estimated to be 28 nA/OD and the limit of detection was calculated to be 0.07 in OD unit. In this work, an immunoassay was prepared by immobilizing anti-HRP antibodies to E.coli cells with autodisplayed Z-domains and the amperometric analysis was demonstrated the quantification of bound HRP. For the demonstration of medical diagnosis, C-reactive protein (CRP) was detected by using electrochemical ELISA based on E.coli with autodisplayed Z-domains.
Details
- ISSN :
- 18777058
- Volume :
- 25
- Database :
- OpenAIRE
- Journal :
- Procedia Engineering
- Accession number :
- edsair.doi.dedup.....9232eda58d09b86844a2a3c9eb1af715
- Full Text :
- https://doi.org/10.1016/j.proeng.2011.12.232