VISAR diagnostic at LIL facility
- Source :
- EPJ Web of Conferences, Vol 59, p 13001 (2013)
- Publication Year :
- 2013
- Publisher :
- EDP Sciences, 2013.
-
Abstract
- A Velocity Interferometer for Any Reflector (VISAR) [1, 2] and a Streaked Optical Pyrometer (SOP) [3] were implemented on the “Ligne integration Laser” (LIL) facility. Spatial resolution as good as 10 μm in the target plane and velocity resolution as good as 0.1 km/s can be achieved. Several campaigns were performed in 2010 involving various experimental setups and physical processes: Boron EOS, Pre-compress H2 with special setup of diamond anvil cell and Shock coalescence. This feedback will be of a great help for the Laser Megajoule facility (LMJ) VISAR design.
Details
- Language :
- English
- Volume :
- 59
- Database :
- OpenAIRE
- Journal :
- EPJ Web of Conferences
- Accession number :
- edsair.doi.dedup.....906fb5d3c3386d596824766558709ab4