Partial breast irradiation with the 1.5 T MR-Linac: First patient treatment and analysis of electron return and stream effects

Authors :
O. Dohm
Vanessa Heinrich
M. Nachbar
N. Weidner
Chiara De-Colle
J. Winter
S. Boeke
Savas Tsitsekidis
David Mönnich
Daniel Zips
Cihan Gani
Daniela Thorwarth
Monica Lo Russo
Lorenzo Livi
Source :
Radiotherapy and Oncology. 145:30-35
Publication Year :
2020
Publisher :
Elsevier BV, 2020.

Abstract

Introduction External beam partial breast irradiation (PBI) provides equal oncological outcomes compared to whole breast irradiation when applied to patients with low risk tumours. Recently, linacs with an integrated magnetic resonance image-guidance system have become clinically available. Here we report the first-in-human PBI performed at the 1.5 T MR-Linac, with a focus on clinical feasibility and investigation of the air electron stream effect (ESE) and the electron return effect (ERE) in the presence of the 1.5 T magnetic field, which might influence the dose on the chin (out-of-field dose, due to the ESE), the skin and the lung/chest wall interface (in-field dose, ERE). Methods A 59 years old patient affected by a 15 mm unifocal grade 1 carcinoma not special type of the right breast staged pT1c pN0 cM0 was planned and treated at Unity 1.5 T MR-Linac. To investigate the ERE and the ESE, an MR-Linac treatment plan was simulated without considering the 1.5 T B field using a research version of Monaco (V. 5.19.03). In vivo dosimetry was performed using Gafchromic® EBT3 films placed on top and underneath a 1 cm bolus which was placed on the patient’s chin. The plans with and without 1.5 T magnetic field were compared in terms of dose to the chin, to the skin and to the interface lung/chest wall. Finally, the dose on the chin measured with the in vivo dosimetry was compared with the dose calculated by Monaco. Results PBI using the 1.5 T MR-Linac was successfully performed with a 7 MV photon 7-beams IMRT step-and-shoot plan. The treatment was well tolerated, the patient developed a slight acute toxicity, i.e. breast skin erythema and breast oedema CTC V.4 grade 1. The plan with 1.5 T magnetic field documented a fractional dose of 0.17 Gy in the chin area (2.6 Gy in 15 fractions), which was reduced to 0.05 Gy (0.75 in 15 fractions) by the presence of 1 cm bolus. The simulated plan without magnetic field showed a dose reduced by 2.3 Gy in the chin area. With the in vivo dosimetry a fractional dose of, respectively, 0.12 Gy and 0.034 Gy on top and underneath the bolus were measured (1.8 and 0.51 Gy in 15 fractions). The plan with 1.5 T magnetic field showed a skin D2 of 40 Gy and a skin V35 of 40.2%, which were reduced to, respectively, 39.7 Gy and 24.9% in the simulation without magnetic field. At the interface lung/chest there were no differences in DVH statistics. Conclusion PBI with the 1.5 T MR-Linac was performed for the first time. ESE is accurately calculated by the treatment planning system, can be effectively reduced with a 1 cm bolus and is comparable to dose of cone beam-CT based position verification. The additional dose caused by ERE is not associated with an increased risk of acute toxicity.

Details

ISSN :
01678140
Volume :
145
Database :
OpenAIRE
Journal :
Radiotherapy and Oncology
Accession number :
edsair.doi.dedup.....8cca5827389c8181d10d9e5ed6c62a90
Full Text :
https://doi.org/10.1016/j.radonc.2019.11.025