Back to Search Start Over

Influence of the dental implant macrogeometry and threads design on primary stability: an in vitro simulation on artificial bone blocks

Authors :
Adriano Piattelli
Francesco De Angelis
Giovanna Iezzi
Luca Comuzzi
Felice Lorusso
Margherita Tumedei
Source :
Computer methods in biomechanics and biomedical engineering. 24(11)
Publication Year :
2021

Abstract

The implant macrogeometry and thread profile represent one of the most important factors for a successful achievement of primary stability during the positioning procedure. The aim of the present investigation was to evaluate the insertion torque (IT), removal torque (RT) and Implant stability Quotient (ISQ) of two different implant macrogeometry and thread profile on solid rigid polyurethane model. Two different implants macrogeometries were tested: K2 (Group I) with 11° angle, 1.17 mm pitch and self-cutting V thread profile and K3 (Group II) implants with 30° angle, 0.71 mm pitch and spyral thread profile. A total of 120 specimens (

Details

ISSN :
14768259
Volume :
24
Issue :
11
Database :
OpenAIRE
Journal :
Computer methods in biomechanics and biomedical engineering
Accession number :