Back to Search Start Over

Comparison of Accuracy of Mitral Valve Area in Mitral Stenosis by Real-Time, Three-Dimensional Echocardiography Versus Two-Dimensional Echocardiography Versus Doppler Pressure Half-Time

Authors :
Mingxing Xie
Tsung O. Cheng
Jing Wang
Xinfang Wang
Qing Lu
Source :
The American Journal of Cardiology. 95:1496-1499
Publication Year :
2005
Publisher :
Elsevier BV, 2005.

Abstract

Mitral valve area (MVA) in 30 patients with mitral stenosis (MS) and 34 normal controls was calculated by real-time, 3-dimensional echocardiography (RT3DE); MVA in patients with MS correlated well with the mitral area determined by 2-dimensional echocardiography (r = 0.98) and by pressure half-time (r = 0.90). MVA in normal controls on RT3DE correlated well with MVA on 2-dimensional echocardiography (r = 0.94) and pressure half-time (r = 0.91). There were significant differences between the orifice areas in patients with MS and normal controls. RT3DE can provide not only the anatomic structure of mitral valve apparatus, but also the optimal plane of the smallest mitral valve orifice, and can thus accurately measure the MVA.

Details

ISSN :
00029149
Volume :
95
Database :
OpenAIRE
Journal :
The American Journal of Cardiology
Accession number :
edsair.doi.dedup.....8518b39ba43a427d3d09d8631441c560
Full Text :
https://doi.org/10.1016/j.amjcard.2005.02.023