Back to Search Start Over

Multicenter validation of the magnetic resonance T2* technique for quantification of pancreatic iron

Authors :
Nicolò Schicchi
Giuseppe Peritore
Emanuele Grassedonio
Paolo Preziosi
Vincenzo Positano
Antonino Vallone
Calogera Gerardi
Alessia Pepe
Ada Riva
Stefania Renne
Laura Pistoia
Gennaro Restaino
Angelo Peluso
Daniele De Marchi
Riccardo Righi
Antonella Meloni
Source :
European radiology. 29(5)
Publication Year :
2018

Abstract

To assess the transferability of the magnetic resonance imaging (MRI) multislice multiecho T2* technique for pancreatic iron overload assessment. Multiecho T2* sequences were installed on ten 1.5-T MRI scanners of the three main vendors. Five healthy subjects (n = 50) were scanned at each site. Five patients with thalassemia (n = 45) were scanned locally at each site and were rescanned at the reference site within 1 month. T2* images were analyzed using a previously validated software and the global pancreatic T2* value was calculated as the mean of T2* values over the head, body, and tail. T2* values of healthy subjects were above 26 ms and showed inter-site homogeneity. The T2* values measured in the MRI sites were comparable to the correspondent values observed in the reference site (12.02 ± 10.20 ms vs 11.98 ± 10.47 ms; p = 0.808), and the correlation coefficient was 0.978 (p

Details

ISSN :
14321084
Volume :
29
Issue :
5
Database :
OpenAIRE
Journal :
European radiology
Accession number :