Back to Search Start Over

Work in progress: a correction for tissue iodine accumulation in videodensitometric measurements of left ventricular ejection fraction

Authors :
Walter W. Peppler
M. S. Van Lysel
P D Myerowitz
C. Mistretta
Dhanani Sp
David K. Swanson
Bruce H. Hasegawa
Source :
Radiology. 147:37-43
Publication Year :
1983
Publisher :
Radiological Society of North America (RSNA), 1983.

Abstract

Left ventricular ejection fractions were determined following eight intravenous injections in three dogs using area/length and digital videodensitometric techniques. Ejection fractions were measured by both techniques for all beats during left ventricular opacification. Even after noniodinated background corrections, tissue iodine accumulation produced large errors in measured iodine content late in the contrast material curve. By using a model for tissue iodine accumulation, an algorithm was developed that appears to reduce these errors. The measured ejection fraction for ten beats following the peak of the contrast material curve declined an average of 1.8% following correction, compared with 14% prior to correction. Following correction for tissue iodine accumulation, correlation between area/length and video-densitometric ejection fractions was 0.94. By correcting for tissue iodine accumulation, much more of the contrast material curve provides usable data for estimating ejection fractions.

Details

ISSN :
15271315 and 00338419
Volume :
147
Database :
OpenAIRE
Journal :
Radiology
Accession number :
edsair.doi.dedup.....83b8a89829f82d76f1e1b03bb8d7cdc6
Full Text :
https://doi.org/10.1148/radiology.147.1.6338564