Back to Search Start Over

A Double-Sideband Continuous-Wave Radar Sensor for Carotid Wall Movement Detection

Authors :
Simone Chicarella
Stefano Pisa
Orlandino Testa
Renato Cicchetti
Emanuele Piuzzi
Erika Pittella
Source :
IEEE Sensors Journal. 18:8162-8171
Publication Year :
2018
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2018.

Abstract

A radar sensor suitable to record carotid wall movements has been developed and tested. The sensor is based on a double-sideband continuous-wave coherent transceiver, operating at 2.45 GHz, and on a bow-tie planar antenna optimized to operate in contact with the neck. Measured transceiver responses on a sliding short load are in good agreement with an analytical model and simulation results. An artificial model of the neck with an embedded pulsatile carotid-artery mockup has been built to test the radar. When the radar antenna is placed in contact with the neck model, an output voltage signal following the carotid diameter variations is measured. The signal amplitude follows the pressure wave imposed on the carotid artery mockup.

Details

ISSN :
23799153 and 1530437X
Volume :
18
Database :
OpenAIRE
Journal :
IEEE Sensors Journal
Accession number :
edsair.doi.dedup.....833559fd2f040c1278055b7a50367528
Full Text :
https://doi.org/10.1109/jsen.2018.2862430