Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber reading

Authors :
J.J.W. Lagendijk
J G M Kok
A H L Aalbers
Bas W. Raaymakers
K Smit
B. Van Asselen
Source :
Physics in Medicine and Biology. 58:5945-5957
Publication Year :
2013
Publisher :
IOP Publishing, 2013.

Abstract

In the UMC Utrecht a prototype MR-linac has been installed. The system consists of a 6 MV Elekta (Crawley, UK) linear accelerator and a 1.5 T Philips (Best, The Netherlands) Achieva MRI system. This paper investigates the feasibility to correct the ionization chamber reading for the magnetic field within the dosimetry calibration method described by Almond et al (1999 Med. Phys. 26 1847-70). Firstly, the feasibility of using an ionization chamber in an MR-linac was assessed by investigating possible influences of the magnetic field on NE2571 Farmer-type ionization chamber characteristics: linearity, repeatability, orientation in the magnetic field; and AAPM TG51 correction factor for voltage polarity and ion recombination. We found that these AAPM correction factors for the NE2571 chamber were not influenced by the magnetic field. Secondly, the influence of the permanent 1.5 T magnetic field on the NE2571 chamber reading was quantified. The reading is influenced by the magnetic field; therefore, a correction factor has been added. For the standardized setup used in this paper, the NE2571 chamber reading increases by 4.9% (± 0.2%) due to the transverse 1.5 T magnetic field. Dosimetry measurements in an MR-linac are feasible, if a setup-specific magnetic field correction factor (P1.5 T) for the charge reading is introduced. For the setup investigated in this paper, the P1.5 T has a value of 0.953.

Details

ISSN :
13616560 and 00319155
Volume :
58
Database :
OpenAIRE
Journal :
Physics in Medicine and Biology
Accession number :
edsair.doi.dedup.....81019df4a149bc66f63567bef3ebd8f9
Full Text :
https://doi.org/10.1088/0031-9155/58/17/5945