Abstract
- BackgroundOur laboratory obtained the ISO 15189 accreditation for the plasmatic HIV-1, HBV and HCV viral load (VL) using the m2000 RealTime™ system, which was recently changed for the platform Panther®. Here, we discuss a strategy for performing method validation/verification very quickly.MethodsWe performed the mandatory (repeatability, internal quality assessment [IQA], measurement uncertainty [MU]) and optional technical verifications for CE/IVD assays using the flexible scope range A. We also performed the mandatory assays for the validation of HIV-1 VL in the cerebrospinal fluid (CSF) using the flexible scope range B. The change was checked by following up on the turnaround time (TAT).ResultsThe coefficient of variation (CV%) for repeatability and IQA complied with the limit of 0.25 log. The MU results ranged from 0.04 to 0.25 log copies or IU/mL. The comparisons of methods showed excellent correlations (R2 = 0.96 for the three parameters) but a delayed centrifugation on HCV VL showed variations of up to 2 log IU/mL. An excellent linearity for HIV-1 in the CSF was obtained from 1.5 to 5 log copies/mL with R2 = 0.99. The TAT increased (84%–98%) in routine usage.ConclusionsThe three Aptima assays are well suited for routine laboratory use and can be integrated within less than 2 weeks in accordance with flexible scope range A. Our data allows us to confidently perform HIV-1 VL in CSF following flexible scope range B. Finally, we provide an organizational guide for flexible scope management in molecular virology within a short time frame.
- Subjects :
- 0301 basic medicine
Hepatitis B virus
Coefficient of variation
[SDV]Life Sciences [q-bio]
030106 microbiology
Clinical Biochemistry
Human immunodeficiency virus (HIV)
Hepacivirus
medicine.disease_cause
Real-Time Polymerase Chain Reaction
Turnaround time
03 medical and health sciences
Time frame
medicine
Humans
Mathematics
Blood-Borne Infections
Scope (project management)
Biochemistry (medical)
Reproducibility of Results
General Medicine
Repeatability
Viral Load
3. Good health
Internal quality
030104 developmental biology
Molecular Diagnostic Techniques
pol Gene Products, Human Immunodeficiency Virus
HIV-1
RNA, Viral
Reagent Kits, Diagnostic
Viral load
Biomedical engineering
Subjects
Details
- Language :
- English
- ISSN :
- 14346621 and 14374331
- Database :
- OpenAIRE
- Journal :
- Clinical Chemistry and Laboratory Medicine, Clinical Chemistry and Laboratory Medicine, De Gruyter, 2020, 58 (2), pp.251-260. ⟨10.1515/cclm-2019-0502⟩
- Accession number :
- edsair.doi.dedup.....80ca5f7fa60e698bfd77407b738b9eea
- Full Text :
- https://doi.org/10.1515/cclm-2019-0502⟩