Drill-mounted video guidance for orthopaedic trauma surgery
- Source :
- J Med Imaging (Bellingham)
- Publication Year :
- 2020
-
Abstract
- Purpose: Percutaneous fracture fixation is a challenging procedure that requires accurate interpretation of fluoroscopic images to insert guidewires through narrow bone corridors. We present a guidance system with a video camera mounted onboard the surgical drill to achieve real-time augmentation of the drill trajectory in fluoroscopy and/or CT. Approach: The camera was mounted on the drill and calibrated with respect to the drill axis. Markers identifiable in both video and fluoroscopy are placed about the surgical field and co-registered by feature correspondences. If available, a preoperative CT can also be co-registered by 3D-2D image registration. Real-time guidance is achieved by virtual overlay of the registered drill axis on fluoroscopy or in CT. Performance was evaluated in terms of target registration error (TRE), conformance within clinically relevant pelvic bone corridors, and runtime. Results: Registration of the drill axis to fluoroscopy demonstrated median TRE of 0.9 mm and 2.0 deg when solved with two views (e.g., anteroposterior and lateral) and five markers visible in both video and fluoroscopy-more than sufficient to provide Kirschner wire (K-wire) conformance within common pelvic bone corridors. Registration accuracy was reduced when solved with a single fluoroscopic view ( TRE = 3.4 mm and 2.7 deg) but was also sufficient for K-wire conformance within pelvic bone corridors. Registration was robust with as few as four markers visible within the field of view. Runtime of the initial implementation allowed fluoroscopy overlay and/or 3D CT navigation with freehand manipulation of the drill up to 10 frames / s . Conclusions: A drill-mounted video guidance system was developed to assist with K-wire placement. Overall workflow is compatible with fluoroscopically guided orthopaedic trauma surgery and does not require markers to be placed in preoperative CT. The initial prototype demonstrates accuracy and runtime that could improve the accuracy of K-wire placement, motivating future work for translation to clinical studies.
- Subjects :
- medicine.medical_specialty
medicine.diagnostic_test
Drill
business.industry
Image-Guided Procedures, Robotic Interventions, and Modeling
ComputingMethodologies_IMAGEPROCESSINGANDCOMPUTERVISION
Image registration
Video camera
030218 nuclear medicine & medical imaging
law.invention
Surgery
03 medical and health sciences
0302 clinical medicine
law
Feature (computer vision)
030220 oncology & carcinogenesis
Fracture fixation
medicine
Fluoroscopy
Radiology, Nuclear Medicine and imaging
Kirschner wire
Guidance system
business
Subjects
Details
- ISSN :
- 23294302
- Volume :
- 8
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- Journal of medical imaging (Bellingham, Wash.)
- Accession number :
- edsair.doi.dedup.....80ac44e8e54aff48a44c964949f05d1f