Real-Time Detection and Alerting for Acute ST-Segment Elevation Myocardial Ischemia Using an Implantable, High-Fidelity, Intracardiac Electrogram Monitoring System With Long-Range Telemetry in an Ambulatory Porcine Model
- Source :
- Journal of the American College of Cardiology. (11):2306-2314
- Publisher :
- American College of Cardiology Foundation. Published by Elsevier Inc.
-
Abstract
- Objectives The purpose of this study was to evaluate feasibility of using real-time, high-fidelity, intracardiac electrogram monitoring from a permanently implantable ischemia detection system (IIDS), with long-range telemetry capability to detect ST-segment shifts associated with acute or subacute coronary occlusion in a porcine model. Background Early identification of coronary occlusion with ST-segment elevation could profoundly accelerate the timing of revascularization and improve clinical outcomes. Methods This paper reports the first investigation using real-time, high-fidelity, intracardiac electrogram monitoring from a permanently IIDS, with long-range telemetry capability. This IIDS was tested in an ambulatory porcine model, with acute coronary occlusion precipitated by stent thrombosis. Two overlapping copper stents were implanted in the left anterior descending (n = 3), the circumflex (n = 3), or the right coronary artery (n = 2) of juvenile farm pigs. Monitoring was carried using telemetry from the IIDS. Results All stented pigs had acute ST-segment elevation event(s) triggering the alerting thresholds of the IIDS. All triggered events were confirmed to be caused by thrombosis of the copper stent(s), and well correlated to infarct age and location. Four of the 8 pigs died from ventricular fibrillation, recorded by the IIDS at a mean time of 70 ± 121 h after ST-segment alert. The sensitivity and specificity of alerting for ST-segment shift, associated with thrombotic coronary occlusion, were 100% and 100%, respectively. Conclusions This study demonstrates the ability of an implantable ischemia detection system to detect ST-segment elevation from coronary occlusion in a porcine model of ST-segment elevation myocardial infarction. ST-segment elevation was sufficient to trigger alerting thresholds in all 3 epicardial coronary distributions. Such a system, with real-time alerting capability, could advance the time frame of reperfusion therapy and potentially prevent, rather than interrupt, acute myocardial infarction in patients with coronary artery disease.
- Subjects :
- medicine.medical_specialty
Swine
medicine.medical_treatment
Myocardial Ischemia
Differential Threshold
Monitoring, Ambulatory
Coronary Disease
Sensitivity and Specificity
Coronary artery disease
Electrocardiography
Reperfusion therapy
Computer Systems
Internal medicine
medicine.artery
medicine
Animals
Telemetry
Myocardial infarction
medicine.diagnostic_test
business.industry
Coronary Thrombosis
Electrodiagnosis
ST elevation
Stent
Equipment Design
Prostheses and Implants
medicine.disease
Coronary occlusion
Right coronary artery
Acute Disease
Ventricular Fibrillation
Cardiology
Feasibility Studies
Stents
Cardiology and Cardiovascular Medicine
business
Subjects
Details
- Language :
- English
- ISSN :
- 07351097
- Issue :
- 11
- Database :
- OpenAIRE
- Journal :
- Journal of the American College of Cardiology
- Accession number :
- edsair.doi.dedup.....7faf6847a648b9c9bef8d84b4aaac883
- Full Text :
- https://doi.org/10.1016/j.jacc.2006.07.058