Back to Search Start Over

Free-breathing, zero-TE MR lung imaging

Authors :
Luigi Landini
Fabio Gibiino
Laura Sacolick
Florian Wiesinger
Anne Menini
Publication Year :
2015

Abstract

The investigation of three-dimensional radial, zero-echo time (TE) imaging for high-resolution, free-breathing magnetic resonance (MR) lung imaging using prospective and retrospective motion correction. Zero-TE was implemented similarly to the rotating-ultra-fast-imaging-sequence, providing 3D, isotropic, radial imaging with proton density contrast. Respiratory motion was addressed using prospective triggering (PT), prospective gating (PG) and retrospective gating (RG) with physiological signals obtained from a respiratory belt and interleaved pencil beam and DC navigators. The methods were demonstrated on four healthy volunteers at 3T. 3D, radial zero-TE imaging with high imaging bandwidth and nominally zero echo-time enables efficient capture of short-lived signals from the lung parenchyma and the vessels. Compared to Cartesian encoding, unaccounted for free-breathing respiration resulted in only benign blurring artifacts confined to the origin of motion. Breath holding froze respiration but achieved only limited image resolution (~1.8 mm, 30 s). PT and PG obtained similar quality expiratory-phase images at 1.2 mm resolution in ~6 min scan time. RG allowed multi-phase imaging in ~15 min, derived from eight individually stored averages. Zero-TE appears to be an attractive pulse sequence for 3D isotropic lung imaging. Prospective and retrospective approaches provide high-quality, free-breathing MR lung imaging within reasonable scan time.

Details

Language :
English
Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....7f8c38c415a22a4e370b896be86459e7