Back to Search Start Over

Development and validation of a quantitative assay for the measurement of miltefosine in human plasma by liquid chromatography-tandem mass spectrometry

Authors :
Michel J.X. Hillebrand
Jos H. Beijnen
Thomas P. C. Dorlo
Hilde Rosing
Peter J. de Vries
Teunis A. Eggelte
Infectious diseases
KIT: Biomedical Research
Source :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences, 865(1-2), 55-62. Elsevier
Publication Year :
2008

Abstract

A sensitive and specific liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for the quantification of miltefosine is presented. A 250 mu L human EDTA plasma aliquot was spiked with miltefosine and extracted by a solid-phase extraction method. Separation was performed on a Gemini C 18 column (150 mm x 2.0 mm I.D., 5 mu m) using an alkaline eluent. Detection was performed by positive ion electrospray ionization followed by triple-quadrupole mass spectrometry. The assay has been validated for miltefosine from 4 to 2000 ng/mL using 250 mu L human EDTA plasma samples. Results from the validation demonstrate that miltefosine can be accurately and precisely quantified in human plasma. At the lowest level, the intra-assay precision was lower than 10.7%, the inter-assay precision was 10.6% and accuracies were between 95.1 and 109%. This assay is successfully used in a clinical pharmacokinetic study with miltefosine. (c) 2008 Elsevier B.V. All rights reserved

Details

Language :
English
ISSN :
15700232
Volume :
865
Issue :
1-2
Database :
OpenAIRE
Journal :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Accession number :
edsair.doi.dedup.....7ed831e4d56d4c78ff1903066e756a71
Full Text :
https://doi.org/10.1016/j.jchromb.2008.02.005